Components | CAS no | Sample [μg/m3] |
Empty chamber [μg/m3] |
Consumption [%] |
---|---|---|---|---|
Methylethylketone (MEK) | 78-93-3 | <1 | 140 | 100% |
Methoxy-2-propanol | 107-98-2 | <1 | 160 | 100% |
Pentanal | 110-62-3 | <1 | 140 | 100% |
Cyclohexanone | 108-94-1 | <1 | 150 | 100% |
Ethyl hexanal | 108-95-2 | <1 | 97 | 100% |
Phenol | 104-76-7 | <1 | 120 | 100% |

ASTEC Gas Line Filters
Enhancing conditions for successful embryo development in your incubator.
The new ASTEC Gas Line Filters are a cost-effective method to eliminate contaminants like VOCs, CACs, microorganisms, aldehydes and particles from incoming CO2, N2 and pre-mixed gas, enhancing conditions for successful embryo development in your incubator.
Six month lifespan allows long term durability and cost effectiveness.
Enhanced Performance: Utilizing powder-form activated carbon, our latest model outperforms the previous granular-form design, ensuring high VOC removal rates. (Refer to Technical Data for detailed performance metrics)
Safe for Uninterrupted Embryo Development
Recognizing the impact of chemicals on embryo development, our Gas Line Filters are manufactured without the use of harmful substances. Furthermore, both the capsule and internal materials are designed to meet safety standards for usage in IVF laboratories, ensuring a secure environment for embryo development.
4 Layers Structure
Technical DATA VOC Test results

Line Connector : VFAS-CNC02(2 pieces)

Holder : VFAS-HLD01
Specification
Material | powder form activated carbon |
---|---|
Filter | Hydrophobic PTEF 0.22 |
External Dimensions (mm) | 180×77.4 |
Sterile | γSterile |
Number/Box | 2/1 |
Recommended Exchange Frequency | Every 6 months |
Safe Usage Pressure | 0.1 – 0.4 Bar |
Maximum Load Pressure | 0.08Mpa(0.03-0.06Mpa recommended) |
Product Usage | Incubators for embryos |
Expiration Date | 3~5 years (in a vacuum), 2 years (normal) |