Minimum quantity Discount
Buy 2 + Get 10% discount
Skip to content

Saturn 5 Active™ Directional Laser System [Quote]

Regular price $0.00

SKU: RI-SAT-ACT

Saturn 5 Active

The Saturn 5 Active is a moveable laser designed for assisted hatching, zona thinning and biopsy techniques in the ART laboratory.

Key Features

  • Fast biopsies - Biopsy Mode provides excellent accuracy while laser drilling along a straight or curved line
  • Minimal laser energy - highest laser power combined with a very short pulse time, which minimizes the sample damage²
  • Intuitive software - RI Viewer software package with streamlined user interface. Optional programmable foot pedal controls software and laser functions
  • Versatile - Compatible with all popular brands of micromanipulators and inverted microscopes

Using Biopsy Mode, users can draw a straight or curved line along a sample, select the number, frequency and size of holes, then ablate along the chosen path without moving the holding pipette. 

Microscope Compatibility

  • Nikon TE200/300, TE2000, Ti, Ti2
  • Leica DMIRB, DMI3000B/4000B/6000B, DMIL, DMi8
  • Olympus IX50/70, IX51/71/81, IX53/73/83
  • Zeiss Axiovert 100/200, Axio Observer

Specifications

Pilot laser  630−650mm spot targeting solid state diode laser – red pilot beam guarantees the position of the invisible ablation laser. It is impossible for the pilot and ablation

lasers to be out of alignment with each other

Objective Custom designed objective for optimum laser transmission, crystal clear imaging and minimal laser pulse times. Tested and proven not to exhibit astigmatism with Saturn 5 Laser Systems. 20x and 40x objectives are available.
Ablation laser

1480nm / 400mW solid state diode laser.

Pulse length range 0.005−2.0ms / 5−2000µs.

Class 1 laser product. Tested and proven not to exhibit thermal lensing

Laser unit dimensions

(WxDxH) 220mm x 180mm x 34mm

Fluorescence compatibility

Compatible with epi-fluorescence on selected microscopes

Operation software

RI Viewer imaging software included, with digital laser targeting

PC system requirements

Operating systems: Windows 10, Windows 8.1, Windows 8, Windows 7 (SP1)

Mains input

100-240VAC, 50-60Hz

 

Back to top

Added to cart