Minimum quantity Discount
Buy 2 + Get 10% discount
Skip to content

Integra 3™ Micromanipulator Micromanipulation System [Quote]

Regular price $0.00

SKU: RI-INT-3

Integra 3™

The Integra 3 has been designed to create the best possible conditions for ICSI and sample manipulation. It ensures gametes and embryos remain at optimum temperature using unique Thermosafe™ technology and through advanced mechanical design, it is ultra-responsive, smooth and easy to use. You can change micropipettes in seconds, transforming the efficiency of your work, and use oil or air syringes for sample control.

Featuring Thermosafe

Thermosafe maintains sample temperature across the dish by passing a gentle stream of warmed air under the Petri dish. The thermal images below, taken with a 40x objective in place on an inverted microscope, display the cooling effect of the objective on the centre of the dish with and without Thermosafe.

Key Features

  • No cold spots – Thermosafe maintains sample temperature with a gentle stream of warm air under the Petri dish
  • Super-fast – Improved rapid micropipette set-up with one-step angle adjustment
  • Practical – Motion sensor LED light underneath the stage for improved visibility when changing objectives
  • Safe – Heating system status indicator light and alarm monitors every second of the process. A log is kept of every alarm triggered
  • Integrated – Large built-in mechanical XY stage plus temperature controllers with easy access buttons and touch screen
  • Accurate – New extra-fine mechanisms for smooth and precise movement of the stage and micropipettes

Microscope Compatibility

  • Nikon - TMD, Diaphot 200/300, TE200/300, TE2000, Ti, Ti2
  • Leica - DMIRB, DMI3000B/4000B/6000B, DMIL, DMi8
  • Olympus - IMT2, IX50/70, IX51/71/81, IX53/73/83
  • Zeiss - Axiovert 40/100/200, Axio Observer, Axio Vert.A1

Specifications

Heating system 

Thermosafe air heating system with heated metal insert: 25mm aperture (standard),16mm aperture (optional), ITO heated glass insert (optional)

Temperature controller 

Accuracy better than ±0.1ºC when calibrated against known reference. Resolution of 0.1ºC. Setpoint temperature range: 30-45ºC

Display 

3.5’’ color LCD touch screen display for temperature control

Connectivity 

USB socket for connection to PC running RI Viewer™ software (optional software)

Syringes 

Choice of SAS (air), SAS-SE (air) and SOS (oil). Compatible with other ART syringes

Weight 

Maximum weight of microscope mounted components (excluding front and rear adaptors): 10.4kg. Actual weight is configuration dependent

Dimensions 

Footprint (not including microscope): width 56cm, depth 38cm

Power supply 

Input: 100-240VAC, 50-60Hz, Max. 1.8A, Class 1.
Output: 12VDC, Max. 11A (132W)

Operating range 

Temperature: 15ºC (59ºF) to 40ºC (104ºF)
Humidity: 15% to 85% RH (non condensing)